PC Automatic Quality Assurance
System for MRI

Features:

  • DICOM Compatible.

  • Graphical Result Screens

  • Summary Report

  • Database File Output Results

  • Provides for Variations in Phantom Configurations and Physical Tests.

  • System Adaptable to the Magphan(tm) Phantom(tm).

  • Monitors System Performance and Service Delivery.

  • Helps Fulfill Regulatory Requirements.

Signal-To-Noise Ratio:
This parameter is calculated by a commonly used equation by determining the quotient of the difference in mean values between a central region and the background region, as divided by the standard deviation of the central region.

Geometric Distortion:
Also referred to as linearity (the ability to image a straight line as a straight line). Influencing this parameter is Bo and x and y gradient linearity. A simplified method of measuring gradient non-linearity can be accomplished by linear determination of the phantom's dimensions.

Uniformity:
Vertical and horizontal profiles are generated through the Phantom's center. The fractional uniformity is calculated as the percentage of the profile pixels within an acceptable range determined by +/-10 of the central mean. The integral uniformity is calculated from the profile data as well.

Spatial Resolution (Pixel Size):
This measurement tracks the degree of electronic temporal sampling of the RF signal in the phase encoding (y) and the readout direction (x), thus reflecting the sampling matrix fineness. For MRI, the system resolution is determined by the aquisition matrix.

Slice Width:
Slice width is derived from gradient induced resonant excitation regions of the magnetic and RF fields. Methodology for this measurement resembles that for CT, whereby the full-width at half maximum (FWHM) of the background corrected ramp profile is equivalent to the experimental slice thickness.


System Requirements:

The system requirements for the IRIS, Inc software packages are as follows:

Pentium Processor
3.5" HD Floppy Drive
DICOM Supported System:Network Interface
Windows 95 Required
Interfaces for Non-DICOM Systems Available